Corneal Pachymetry: Modalities and Instruments

Corneal Pachymetry:  Modalities and Instruments

Michael R. Feilmeier, MD
Bascom Palmer Eye Institute

Corneal pachymetry is the measurement of corneal thickness. Pachymetry was traditionally used to gauge the functional status of the corneal endothelial cell layer. More recently, with the emergence of refractive surgical techniques, corneal pachymetry is necessary to determine suitable candidates for ablation procedures. Furthermore, the identification of central corneal thickness (CCT) as an independent indicator of glaucoma risk by the Ocular Hypertensive Treatment Study (OHTS) has made corneal pachymetry a routine part of the ophthalmic evaluation. Pachymetry is an important part of the evaluation and management of ocular hypertension and glaucoma.

Several techniques are available to reliably and reproducibly measure corneal thickness. The methodologies used in these techniques are based on either ultrasonic or optical principles. While each of the methods have a peculiarity of their own, all have been described as reliable. Keep in mind that systematic differences exist between the different techniques and result in different values. As a result, the measurements cannot simply be substituted between the different modalities.

Traditional ultrasound pachymetry (10-20 MHz) offers the advantages of portability and relative ease of use. These dry contact systems are simple, portable, and cost-effective. However, the accuracy of this technique is dependent on the perpendicularity of the probe's application to the cornea and the reproducibility relies on precise probe placement. Ultrasound biomicroscopy (50 MHz) and very-high-frequency ultrasound (70 MHz) have the disadvantage of requiring a water bath, but allow for determination of corneal sublayer detail and pachymetry.

Some ultrasound pachymeters contain additional features beyond simply providing a CCT measurement. For example, the Reichert IOPac features a built-in glaucoma risk calculator. This tool helps physicians identify a patient’s 5-year risk of progression in glaucoma, based on the 6 primary risk factors identified by the OHTS study: Age, Race, IOP, Patter Standard Deviation, Cup to Disk ratio, and CCT.

Optical slit lamp techniques measure corneal thickness using a device mounted to the slit lamp through which the observer aligns the anterior surface and endothelial surface of the cornea through image doubling. The corneal thickness is then estimated using an equation based on assumptions of the refractive index and anterior radius of curvature of the cornea. These variables, along with the examiner-dependent nature of the technique are the major disadvantages of this modality.

Specular microscopy is a contact based method that records the adjustment required in the focal plane of the instrument. The SP-2000P specular microscope (Topcon Corp.) is a noncontact optical instrument that provides pachymetry and specular microscopy simultaneously.

Orbscan is a noncontact optical scanning-slit instrument that provides pachymetry in addition to topographic analysis. It has been reported that this technique underestimates corneal thickness in both corneas that have undergone refractive surgery and in corneas with haze.

Optical coherence tomography (OCT) and optical low-coherence reflectometry (OLCR) are noncontact techniques that acquire pachymetry measurements based on optical interferometry. These techniques are able to discern sublayer detail and pachymetry and may be able to perform measurements on corneas with pathology that do not allow measurement by other techniques. OCLR is advantageous in that it provides continuous measurement and is useful in obtaining pachymetry measurements during corneal ablation procedures.

Confocal microscopy is a contact technique that acquires measurements by through focusing a confocal microscope through the thickness of the cornea. It has the advantage of displaying and allowing measurement of detailed sublayer cellular structure. It is also a useful adjunct in the identification of corneal microbial pathology. Its use is significantly limited in the setting of corneal opacities and the technique is inherently timely in terms of both data acquisition and interpretation.

The Pentacam Scheimpflug Camera is a noncontact technique that uses a rotating Scheimpflug camera to rapidly capture images of the anterior segment of the eye. This technique allows measurement of corneal thickness from limbus to limbus.

Laser Doppler interferometry is a noncontact technique that uses a dual-beam infrared laser Doppler interferometry to measure corneal thickness. There is limited data available in the literature regarding this technique.